Lee, Park, Baek, Kim, Lee, and Ryou: Undeflatable balloon guide catheter (BGC) during endovascular procedure: Rescue strategy

Abstract

The use of a balloon guide catheter (BGC) in the endovascular management of acute ischemic stroke is known to improve the efficacy and efficiency of the procedure by reducing the risk of distal embolization. During the procedure, the balloon of the catheter causes a temporary arrest of cerebral blood flow. However, failure of the balloon to deflate during the BGC procedure can result in catastrophic complications, including aggravated hypoxic damage.
This paper aims to share the resolution and methodological analysis of our experience with BGC balloon deflation failure, which was confirmed by a reproducible experiment under similar conditions.

INTRODUCTION

Following the recent publication of five randomized trials (MR CLEAN, SWIFT PRIME, EXTEND-IA, ESCAPE, REVASCAT, THRACE) [8,20], mechanical thrombectomy, along with medical management, was established as the standard treatment for acute ischemic stroke caused by large vessel occlusion. Delays in performing mechanical thrombectomy for the purposes of awaiting results of intravenous alteplase for treatment of acute cerebral infarction are discouraged in the 2019 AHA/ASA guideline [8], as are recommendations to perform immediate mechanical thrombectomy on patients who visit the hospital within 6 h of the onset of symptoms and do not meet the exclusion criteria [8,20]. If the eligibility criteria for DAWN (within 6-16 h) or DEFUSE 3 (within 16-24 h) trials are met [1,17,20], mechanical thrombectomy is possible even in cases in which more than 6 h have passed since the onset of symptoms or the last normal time is unknown.
In the treatment of acute ischemic stroke, various mechanical thrombectomy strategies, such as stent retrieval technique and direct aspiration thrombectomy [20,27], are used in conjunction with novel devices such as balloon guide catheters (BGC) or large-bore distal-access catheters [3,10,16,31] for increased efficacy and efficiency [20]. The BGC, in particular, is known to lower the risk of distal embolization during the procedure and improve flow reversal effects by the deployment of a balloon, which causes the arrest of cerebral blood flow [7,21,28], Such an advantage has been reported in two recent clinical studies to increase recanalization rate and improve neurological outcome [6,16,27,28].
However, failure of the balloon to deflate during the BGC procedure can result in catastrophic complications, including aggravated hypoxic damage. This paper aims to share the resolution and methodological analysis of our experience with BGC balloon deflation failure, which was confirmed by a reproducible experiment under similar conditions.

CASE PRESENTATION

A 104-year-old female patient presented to the emergency department with a chief complaint of altered consciousness and right hemiparesis. She was apparently normal 5 h prior to the presentation. No cerebral hemorrhage was detected on non-contrast brain computed tomography (CT). The patient had an Alberta Stroke Program Early CT Score (ASPECTS) of seven and a National Institutes of Health Stroke Scale (NIHSS) score of 15. An occlusion in the left proximal M1 segment was detected (Fig. 1A) on CT angiography of the brain, and mechanical thrombectomy was performed immediately (Fig. 2A).

Procedure for mechanical thrombolysis

The procedure was performed under local anesthesia. Flowgate BGC (8-Fr, Stryker Neurovascular, Fremont, CA, USA) (Fig. 3A) was inserted through the femoral sheath and placed proximally near the left internal carotid artery (ICA). During the procedure the balloon of BGC was inflated with 1:1 mixed contrast solution. Left ICA angiography revealed occlusion of the left proximal M1 segment, and stent retrieval was performed using the Trevo XP ProVue stent (4 mm×20 mm, Stryker Neurovascular, Fremont, CA, USA). Although the left middle cerebral artery (MCA) was recanalized at the time the stent was deployed, several re-occlusions occurred following stent retrieval. Additional attempts were made to recanalize the MCA using the ADAPT [12,23] and Solumbra techniques [29,30] with the AXS Catalyst 6 Distal Access Catheter (Stryker Neurovascular, Fremont, CA, USA), but the outcomes remained the same. During the procedure, the balloon of the BGC failed to deflate (Fig. 2B). Although aspiration with a syringe was attempted several times to de-balloon the BGC, it failed to de-balloon and intracerebral blood flow arrest persisted with the balloon fixed to the proximal ICA.

Troubleshooting

To enable immediate de-ballooning, a transverse incision was made along the proximal part of the BGC (Fig. 3B). Because there is a channel embedded within the wall of the BGC that allows the balloon to inflate (Fig. 3C), it was assumed that removing the catheter would expose the inflation channel or space, releasing the contrast solution in the balloon and ultimately de-ballooning it. However, deflation once again failed, despite the transverse excision of the proximal portion of the BGC. An intraluminal approach by passing a cutting balloon through the BGC was attempted to deflate the balloon. A cutting balloon is a device with three to four microsurgical blades attached to the balloon surface (Fig. 3D) [5,25] that is designed to create microincisions in surrounding structures when inflated (Fig. 3E).
A 5-mm peripheral cutting balloon (5.0 mm×2.0 cm, Boston Scientific Corporation, Marlborough, MA, USA) with a diameter slightly smaller than the proximal ICA was used to introduce the device into the BGC. The cutting balloon was inflated after being inserted and placed near the BGC balloon, and the process was monitored in real-time using fluoroscopy. While the cutting balloon was being inflated, the balloon of the BGC remained inflated. However, a few seconds after the cutting balloon was inflated, the balloon of the BGC gradually decreased in size and eventually became completely deflated. As a result, the BGC was successfully removed from the ICA and the left ICA flow was recovered from a state of flow arrest (Fig. 2C). All it takes 15 minutes to bail out and there was no interval change in the patient’s neurologic condition.

DISCUSSION

Endovascular treatment has recently been established as an essential treatment for acute ischemic stroke [8,20]. BGC, in particular, has been widely used to improve the success rates of recanalization and shorten procedure time during stent retrieval thrombectomy [16,28].
Although there are various complications associated with endovascular procedures, there have been reports of coronary artery intervention for complications related to deflation failure [13,24,26]. The EuroPCR group has proposed a number of rescue therapeutic options in cases of non-deflatable coronary balloons (www.pcronline.com) [19].
Brinjikji et al. reported groin complications (groin or retroperitoneal hematoma, femoral artery bleeding, etc.) caused by large groin sheaths [6], whereas catheter instability, carotid artery dissection, and difficult aortic arch have been identified as potential complications [9]. However, there have been no reports of balloon deflation failures or discussions regarding potential solutions in the case of balloon deflation failure as observed in this case.
Balloon deflation failure is a very rare complication of BGC that occurred in this case, and the potential causes and solutions were analyzed. First, the composition ratio of the contrast solution used in the balloon may influence deflation failure. A 1:1 dilution of the contrast has been suggested as an optimal concentration considering visibility and balloon deflation time [11,14]. The contrast solution used in this case was also 1:1. Second, potential mechanical factors may also influence the deflation of the balloon. However, the possibility of mechanical error was excluded in this case, as the deflation failure occurred after several trials of successful ballooning and de-ballooning. The anatomical factors of the patient may also influence the deflation of the BGC. The patient in this case was an adult 104-year-old adult with a very steep vasculature entry angle and tortuous morphology (Fig. 1B). Her left ICA had a coiling morphology with a circular configuration, and her aorta had a vertical distance between the innominate artery and the top of the aortic arch that was more than twice the diameter of the left common carotid artery, which is considered unfavorable for an endovascular approach procedure (Fig. 1C, D) [2,15]. Even if the BGC was safely positioned at the proximal ICA before the circular segment, the position of the BGC may have shifted incidentally during the repetitive ballooning and de-ballooning process and stent retrieval. As a result, structural changes or damage to the inflation lumen of the BGC, such as torsion or kinking, could have occurred, leading to deflation failure.
Given the mechanism by which the balloon deflated after several attempts, it was surprising why the excision of the proximal portion of FlowGateTM did not cause the balloon to deflate. The FlowGateTM structure includes a lumen-shaped channel (inflation lumen) that is used to inflate the BGC balloon through an external injection of contrast solution (Fig. 3C). Because this lumen is embedded within the catheter wall, excision of the catheter should result in immediate drainage of the contrast solution and subsequent deflation of the balloon, though this did not happen in this study. Furthermore, given that deflation occurred immediately after the introduction and inflation of the cutting balloon, it is unclear whether the deflation was caused by the cutting balloon based on the profiles of the BGC and cutting balloon.
The cutting balloon was first invented by Dr. Peter Barath and was mainly used for the treatment of coronary artery diseases [4]. Unlike conventional balloon catheters, cutting balloons have three to four 0.005-inch atherotomes (razor blades) on the balloon surface (Fig. 3D) [4,5]. When the cutting balloon is inflated, the blades are designed to disrupt the fibroelastic continuity of the neointimal hyperplasia by making sharp transverse microincisions (Fig. 3E) [4,18,22]. Although deflation was attempted by inserting the cutting balloon into the BGC, it was impossible to deflate the balloon in the BGC with the 0.005-inch blade, due to the structure of the BGC wall, which requires a blade of at least 0.007 inches (Fig. 3C).
To simulate the situation, a FlowGateTM was inserted into a rubber tube with the thickness of a blood vessel and ballooned with a 1:1 contrast solution. The cutting balloon was then inflated and monitored in real-time using fluoroscopy, but the balloon in the BGC did not deflate. The BGC balloon did not rupture despite several attempts with different positions of the cutting balloon and the application of sufficient pressure to cause the cutting balloon to rupture the BGC balloon (Fig. 4A, B).
The proximal portion of the BGC was then excised. Once a straight BGC was prepared and inflated in the aforementioned rubber tube, followed by excision of the proximal portion of the catheter, solution leakage occurred in the excised portion of the BGC, resulting in complete deflation within 30 s. The time was then recorded, while the BGC was used to create flexures ranging from 0 to 90°. It was discovered that at a 90° flexure, the balloon had only partially deflated even after 5 min, in contrast to the 30 s it took for the balloon to deflate at a 0° flexure (Fig. 4C, D).
In conclusion, the findings show that the cutting balloon does not cause damage to the BGC wall and that deflation is possible by excising a proximal portion of the BGC. However, it was demonstrated that in the presence of flexure in the BGC, the deflation time may be delayed, or only partial deflation may occur. Kinks in the catheter were found to be present in cases like this, where deflation failed to occur even after the proximal portion of the BGC was excised. As a result, inserting a stiff device into the interior of the BGC to straighten the collapsed catheter lumen caused by the kinking may be required as an additional procedure.

CONCLUSIONS

The BGC balloon deflation failure observed in this case is an extremely rare and unknown complication in the neurovascular field that can lead to fatal and catastrophic outcomes if not addressed immediately. Although rare, if the balloon fails to deflate during a BGC procedure, knowing how to restore luminal patency with a stiffer device and excising the BGC will help to successfully deal with this rare complication.

NOTES

Disclosure

The authors report no conflict of interest concerning the materials or methods used in this study or the findings specified in this paper.

Fig. 1.
The anatomic features are known to be unfavorable for endovascular procedures. (A) Shows computed tomographic angiography of the brain (white arrow) shows flow arrest caused by left middle cerebral artery occlusion. (B) Shows the aortic arch and the curvature of the internal carotid artery (ICA) (head of the white arrow) in the patient studied. (C) Shows the “coiling” feature of proximal ICA with extreme tortuosity, and (D) shows the vertical distance between the innominate artery and the top of the aortic arch, which was greater than twice the diameter of the left common carotid artery [2].
jcen-2022-e2021-10-005f1.jpg
Fig. 2.
Cerebral angiography in the patient studied. (A) Shows the anteroposterior view of cerebral angiography after stent deployment for stent retrieval. Because of the recanalization of the left middle cerebral artery, distal flow is visible. However, re-occlusion occurred during stent retrieval. (B) Shows the lateral view of the balloon guide catheter (BGC) in the proximal part of the coiling segment in the left ICA. The sudden deflation of the BGC failed, and there was prolonged arrest of intracerebral flow. (C) Successful deflation was achieved, and intracerebral flow restored following a series of processes involving the excision of a proximal portion of the BGC and the passage of cutting balloon through the BGC.
jcen-2022-e2021-10-005f2.jpg
Fig. 3.
The structural feature of the balloon guide catheter (BGC) and cutting balloon. (A) Is an inflated BGC; it shows leakage of the contrast media from the two white dots (white arrows) above the region in which the balloon is inflated. (B) Shows the excised surface of the BGC, which is relatively clean, and the structure has not been crushed or deformed. (C) Shows the cross-sectional structure of the excised BGC. The balloon of the BGC is located outside the 0.022-inch catheter wall, which consists of an outer shaft, stainless metal braid and inner shaft with thickness of 0.0025-, 0.0015- and 0.0055-inches, respectively. Because, the thickness of the catheter from the lumen to the inflation lumen is 0.007 inches, the 0.005-inch blade of the cutting balloon does not reach the balloon of the BGC, or the inflation lumen. (D) Shows three to four razor blades (at the heads of the black arrows) present on the surface of the cutting balloon. (E) Shows a cross-section of the transverse microincision made by the cutting balloon on the vessel lumen (black arrow) [4]. The BGC wall was expected to be damaged by the blades of the cutting balloon, resulting in the deflation of the balloon.
jcen-2022-e2021-10-005f3.jpg
Fig. 4.
The methodological analysis of our experience with balloon guide catheter (BGC) balloon deflation failure, which was confirmed by a reproducible experiment under similar conditions. (A) Shows the introduction of the cutting balloon in the inflated BGC. (B) No deflation was observed in the BGC balloon, although sufficient pressure was applied for the cutting balloon to rupture the BGC. The contrast media is leaking due to a rupture in the cutting balloon (white arrow). (C) Shows the BGC inflated inside a rubber tube. The balloon did not deflate despite the excision of the proximal portion of the BGC. This may be attributed to the luminal collapse caused by the flexure in the catheter. (D) After excising the proximal portion of the BGC and straightening the kink in the flexure of the catheter, luminal patency is obtained, and deflation occurs automatically within 30 s.
jcen-2022-e2021-10-005f4.jpg

REFERENCES

1. Albers GW, Marks MP, Kemp S, Christensen S, Tsai JP, Ortega-Gutierrez S, et al. Thrombectomy for stroke at 6 to 16 hours with selection by perfusion imaging. N Engl J Med. 2018 Feb;378(8):708-18.
crossref pmid pmc
2. Alverne FJAM, Lima FO, Rocha FA, Bandeira DA, Lucena AFP, Silva HC, et al. Unfavorable vascular anatomy during endovascular treatment of stroke: challenges and bailout strategies. J Stroke. 2020 May;22(2):185-202.
crossref pmid pmc pdf
3. Baek JH, Kim BM, Kang DH, Heo JH, Nam HS, Kim YD, et al. Balloon guide catheter is beneficial in endovascular treatment regardless of mechanical recanalization modality. Stroke. 2019 Jun;50(6):1490-6.
crossref pmid
4. Barath P, Fishbein MC, Vari S, Forrester JS. Cutting balloon: a novel approach to percutaneous angioplasty. The American Journal of Cardiology. 1991 68(11):1249-52.
crossref pmid
5. Bosiers M, Deloose K, Verbist J, Peeters P. Scoring and cutting balloons for SFA disease treatment. Endovascular Today. 2008 34-7.
6. Brinjikji W, Starke RM, Murad MH, Fiorella D, Pereira VM, Goyal M, et al. Impact of balloon guide catheter on technical and clinical outcomes: a systematic review and meta-analysis. J Neurointerv Surg. 2018 Apr;10(4):335-9.
crossref pmid
7. Chueh JY, Kang DH, Kim BM, Gounis MJ. Role of balloon guide catheter in modern endovascular thrombectomy. J Korean Neurosurg Soc. 2020 Jan;63(1):14-25.
crossref pmid pmc pdf
8. Hong KS, Ko SB, Yu KH, Jung C, Park SQ, Kim BM, et al. Update of the Korean clinical practice guidelines for endovascular recanalization therapy in patients with acute ischemic stroke. J Stroke. 2016 Jan;18(1):102-13.
crossref pmid pmc pdf
9. Jeong DE, Kim JW, Kim BM, Hwang W, Kim DJ. Impact of balloon-guiding catheter location on recanalization in patients with acute stroke treated by mechanical thrombectomy. AJNR Am J Neuroradiol. 2019 May;40(5):840-4.
crossref pmid pmc
10. Kang DH, Kim BM, Heo JH, Nam HS, Kim YD, Hwang YH, et al. Effect of balloon guide catheter utilization on contact aspiration thrombectomy. J Neurosurg. 2018 Nov;1-7.
crossref
11. Kim MT, Park JH, Shin JH, Kim N, Kim SD, Tsauo J, et al. Influence of contrast agent dilution on ballon deflation time and visibility during tracheal balloon dilation: a 3D printed phantom study. Cardiovasc Intervent Radiol. 2017 Feb;40(2):285-90.
crossref pmid pdf
12. Lapergue B, Blanc R, Guedin P, Decroix JP, Labreuche J, Preda C, et al. A direct aspiration, first pass technique (ADAPT) versus stent retrievers for acute stroke therapy: an observational comparative study. AJNR Am J Neuroradiol. 2016 Oct;37(10):1860-5.
crossref pmid pmc
13. Leibundgut G, Degen C, Riede F. Transcutaneous puncture of an undeflatable coronary angioplasty balloon catheter. Case Rep Cardiol. 2018 Sep;2018:6252809.
crossref pmid pmc pdf
14. Mogabgab O, Patel VG, Michael TT, Kotsia A, Christopoulos G, Banerjee S, et al. Impact of contrast agent viscosity on coronary balloon deflation times: bench testing results. J Interv Cardiol. 2014 Apr;27(2):177-81.
crossref pmid pmc pdf
15. Nagata T, Masumoto K, Hayashi Y, Watanabe Y, Kato Y, Katou F. Three-dimensional computed tomographic analysis of variations of the carotid artery. J Craniomaxillofac Surg. 2016 Jun;44(6):734-42.
crossref pmid
16. Nguyen TN, Malisch T, Castonguay AC, Gupta R, Sun CH, Martin CO, et al. Balloon guide catheter improves revascularization and clinical outcomes with the Solitaire device: analysis of the North American Solitaire Acute Stroke Registry. Stroke. 2014 Jan;45(1):141-5.
pmid
17. Nogueira RG, Jadhav AP, Haussen DC, Bonafe A, Budzik RF, Bhuva P, et al. Thrombectomy 6 to 24 hours after stroke with a mismatch between deficit and infarct. N Engl J Med. 2018 Jan;378(1):11-21.
pmid
18. Park GS, Ahn TH, Son MS, Shon JW, Ryu ES, Jin DK, et al. Clinical experience of cutting balloon angioplasty for in stent restenosis. Korean Circulation J. 2002 32(4):317-21.
crossref
19. Pasamanickam K, Rajadurai J, Choo G-H. A new technique to puncture and remove a PTCA balloon that fails to deflate. Int J Angiol. 2020.
crossref
20. Powers WJ, Rabinstein AA, Ackerson T, Adeoye OM, Bambakidis NC, Becker K, et al. Guidelines for the early management of patients with acute ischemic stroke: 2019 update to the 2018 guidelines for the early management of acute ischemic stroke: a guideline for healthcare professionals from the American Heart Association/American Stroke Association. Stroke. 2019 Dec;50(12):e344-418.
crossref pmid
21. Schonfeld MH, Kabiri R, Kniep HC, Meyer L, McDonough R, Sedlacik J, et al. Effect of balloon guide catheter utilization on the incidence of sub-angiographic peripheral emboli on high-resolution DWI after thrombectomy: a prospective observational study. Front Neurol. 2020 May;11:386.
crossref pmid pmc
22. Song HH, Kim KT, Chung SK, Kim YO, Yoon SA. Cutting balloon angioplasty for resistant venous stenoses of Brescia-Cimino fistulas. J Vasc Interv Radiol. 2004 Dec;15(12):1463-7.
crossref pmid
23. Spiotta AM, Fargen KM, Chaudry I, Turner RD, Turk AS. ADAPT: a direct aspiration first pass technique. Endovascular Today. 2016 15(2.
24. Takama T, Ito Y, Ishimori H, Tsukahara R, Muramatsu T. Failure of a balloon to deflate during post dilatation in a coronary artery. Cardiovasc Interv Ther. 2015 Jan;30(1):57-60.
crossref pmid pdf
25. Takano Y, Currier JW, Yeatman LA, Kobashigawa JA, Rogers AD, Cianfichi LJ, et al. Cutting balloon angioplasty for cardiac transplant vasculopathy. J Heart Lung Transplant. 2002 Aug;21(8):910-3.
crossref pmid
26. Trivedi R. Double jeopardy: failure to deflate stent balloon in rescue angioplasty. Interv Med Appl Sci. 2019 Jun;11(2):128-30.
crossref pmid pmc
27. Turc G, Bhogal P, Fischer U, Khatri P, Lobotesis K, Mazighi M, et al. European Stroke Organisation (ESO) - European Society for Minimally Invasive Neurological Therapy (ESMINT) guidelines on mechanical thrombectomy in acute ischemic stroke. J Neurointerv Surg. 2019 Jun;11(6):535-8.
crossref pmid
28. Velasco A, Buerke B, Stracke CP, Berkemeyer S, Mosimann PJ, Schwindt W, et al. Comparison of a balloon guide catheter and a non-balloon guide catheter for mechanical thrombectomy. Radiology. 2016 Jul;280(1):169-76.
crossref pmid
29. Wong J, Telischak N, Heit J, Moraff A, Dodd R, Do H, et al. Acute stroke intervention for large vessel occlusion with combined stent retreival and suction thrombectomy (solumbra technique): a retrospective analysis of 85 patients. J Neurointerv Surg. 2016 8:A86.
30. Yi HJ, Sung JH, Lee DH, Song SY. Effectiveness and technical considerations of solitaire platinum 4×40 mm stent retriever in mechanical thrombectomy with solumbra technique. J Korean Neurosurg Soc. 2020 Jan;64(1):30-8.
crossref pmid pmc pdf
31. Zaidat OO, Mueller-Kronast NH, Hassan AE, Haussen DC, Jadhav AP, Froehler MT, et al. Impact of balloon guide catheter use on clinical and angiographic outcomes in the STRATIS stroke thrombectomy registry. Stroke. 2019 Mar;50(3):697-704.
pmid
TOOLS
METRICS Graph View
  • 1 Crossref
  •  0 Scopus
  • 2,152 View
  • 50 Download
ORCID iDs

Taejoon Park
https://orcid.org/0000-0001-7509-7060

Related articles


ABOUT
BROWSE ARTICLES
EDITORIAL POLICY
FOR CONTRIBUTORS
Editorial Office
The Journal of Cerebrovascular and Endovascular Neurosurgery (JCEN), Department of Neurosurgery, Wonkwang University
School of Medicine and Hospital, 895, Muwang-ro, Iksan-si, Jeollabuk-do 54538, Korea
Tel: +82-2-2279-9560    Fax: +82-2-2279-9561    E-mail: editor.jcen@the-jcen.org                

Copyright © 2024 by Korean Society of Cerebrovascular Surgeons and Korean NeuroEndovascular Society.

Developed in M2PI

Close layer
prev next