J Cerebrovasc Endovasc Neurosurg > Volume 27(4); 2025 > Article
Alkhawaldeh, Aljabali, Moawad, Abualhayjaa, and Alsalhi: Adverse events and failures with Cerenovus CEREGLIDE 71 Intermediate Catheter: An analytic review of the FDA MAUDE database

Abstract

Objective

To assess the safety and performance of the Cereglide 71 aspiration catheter, launched in February 2024, by analyzing adverse events from the Manufacturer and User Facility Device Experience (MAUDE) database.

Methods

Reports from the MAUDE database (February 2024 onward) were analyzed using Microsoft Excel and Jamovi, with 36 reports reviewed—27 device-related and 14 patient-related.

Results

The most common device-related issues were cracks (8 occurrences), material bending (6), and deformation/stretching (5). Adverse events included thromboembolism (4 cases), intracranial hemorrhage (2), carotid-cavernous fistula (2), and cerebral vasospasm (2). The most common indication was middle cerebral artery stroke (9 cases).

Conclusions

This study highlights the importance of post-market surveillance to improve patient outcomes, identifying key complications that require careful consideration in clinical practice. This is the first report on complications related to the Cereglide 71 aspiration catheter from the FDA MAUDE database.

INTRODUCTION

The treatment of acute ischaemic stroke has been revolutionised by mechanical thrombectomy (MT), especially for patients with large vessel occlusions (LVO). Aspiration thrombectomy, particularly with large-bore aspiration catheters, has become a very attractive substitute for stent retrievers among the endovascular procedures that are currently available. Aspiration catheters, which frequently call for fewer device passes and shorter procedure times, have the potential to enable faster, easier, and just as effective revascularisation [4,25]. With successful recanalisation rates ranging from 89% to 96% in studies involving devices like the REACT and Sofia catheters, the A Direct Aspiration First Pass Technique (ADAPT) has gained popularity in clinical practice [16,19]. Better navigation, clot engagement, and clinical outcomes have been facilitated by advancements in catheter design, such as bevelled tips, larger inner diameters, and reinforced delivery systems [12,30].
An estimated 7.8 million new cases of ischaemic stroke were reported worldwide in 2021, indicating that the disease’s burden is still significant [31]. Due to ageing populations and enduring vascular risk factors, absolute case numbers are still increasing even though age-standardized incidence rates have decreased in some areas [7,13]. According to estimates, 10-15% of patients who have an acute ischaemic stroke qualify for MT, and demographic changes are expected to result in a 40% increase in treatment demand by 2040 [20,22]. Therefore, public health priorities include increasing access to MT and improving procedural efficiency.
With comparatively comparable rates of TICI 2b/3 reperfusion, functional independence, and safety outcomes, numerous clinical trials and meta-analyses have assessed the relative effectiveness of ADAPT and stent retrievers [21]. However, some studies have linked ADAPT to a lower incidence of distal emboli and faster recanalisation [14]. The decision between aspiration and stent retriever techniques is frequently impacted by anatomical complexity, clot characteristics, and operator preference; however, randomised evidence is still scarce [28,29].
There are still significant gaps in the post-market safety data for more recent aspiration catheters, despite encouraging outcomes in early-phase trials and observational studies. Rare or delayed complications might go unnoticed in controlled settings, even though registries and trials like ASTER and COMPASS offer insights into efficacy [24]. In addition to indicating high recanalisation rates, early assessments of large-bore catheters such as React 71, Sofia Plus, and CEREGLIDE 71 also raise concerns regarding complications like intracranial haemorrhage, thromboembolism, and catheter tip breakage [27]. Furthermore, the generalisability of many studies is diminished by their small sample sizes, brief follow-up times, and selective populations.
Post-market surveillance systems, like the FDA’s Manufacturer and User Facility Device Experience (MAUDE) database, are essential in this regard. MAUDE provides early warning indicators for device-related issues by gathering real-world safety data across a range of environments and operator skill levels, in contrast to clinical trials [5]. Underreporting, non-standardized data fields, and a lack of denominator data are some of MAUDE’s significant drawbacks, which make it difficult to determine actual event rates or conduct systematic device comparisons [2]. However, with careful interpretation, it continues to be a useful supplement to registry-based and pre-market research, particularly for novel or developing devices.
In February 2024, the Cereglide 71 catheter launched with TruCourse Technology to enhance flexibility, navigation, and clot aspiration. It supports direct aspiration and can be used with stent retrievers for versatile thrombectomy. Characterising the post-market safety profile of aspiration devices, such as the CEREGLIDE 71 catheter, is urgently needed due to the rapid introduction of newer aspiration devices and the widespread use of aspiration thrombectomy. Procedures can be improved, device design can be improved, and patient safety can be improved by routinely understanding device-related malfunctions and related patient complications. Early post-market surveillance is both necessary and timely because there are currently no systematic analyses of real-world complications associated with the CEREGLIDE 71 catheter in the literature.
By examining adverse events recorded in the FDA MAUDE database after the Cerenovus CEREGLIDE 71 Intermediate Catheter was introduced to the market in 2024, this study seeks to assess the safety and effectiveness of the device. We compare our results with published clinical data, give a descriptive overview of complications related to both patients and devices, and identify important areas for further safety monitoring and device optimisation.

MATERIALS AND METHODS

This study used the MAUDE database, maintained by the FDA, to monitor adverse events related to medical devices approved for use in the United States. Manufacturers, importers, and device-user facilities are required to report to the MAUDE database, while healthcare professionals, patients, and consumers can do so voluntarily. Since the MAUDE database contains de-identified data and is publicly accessible, ethical approval was not necessary.
The initial search for adverse event reports related to the Cereglide 71 Intermediate Catheter involved using the keywords “Cereglide 71,” “Cereglide 071,” and “Cereglide catheter.” Reports from the FDA MAUDE database began appearing in February 2024. The device fault codes “NRY” (catheter malfunction) and “QJP” (device difficult to position), as well as the company name “Cerenovus,” were used in both basic and advanced search queries.
Initially, 118 reports were identified. After review, 74 were excluded because they concerned different devices, and 8 were excluded due to insufficient information or duplicates.
This left 44 reports specifically about the Cereglide 71 Intermediate Catheter. Out of these, 36 met the inclusion criteria and were included in the final analysis (Fig. 1). Two authors (IMK, HKA) reviewed the report reference numbers and event descriptions to identify duplicates and similar cases. If a report mentioned more than one patient, the cases were separated into individual reports. A third author (MHEM) reviewed and verified the results.
The reports were then filtered to distinguish between patient-related (PR) and device-related (DR) events. Data filtering, organization, and analysis were conducted using Microsoft Excel 2021 and Jamovi, with descriptive statistics applied to evaluate reporting frequency.

RESULTS

During the study period, a total of 36 medical device reports were collected after screening from all reports submitted. These included 27 device-related reports with 47 problems and 14 patient-related adverse events. Six reports involved both patient and device problems, and 14 reports contained multiple device-related events.
Patients experienced various adverse events, such as four cases of thromboembolism, two intracranial hemorrhages (one subarachnoid hemorrhage [SAH] and one acute intracranial hemorrhage [aICH]), two carotid-cavernous fistulas (CCF), two episodes of cerebral vasospasm affecting the internal carotid artery (ICA), and vascular dissections involving the ICA and right vertebral artery. Additionally, one patient suffered an ischemic stroke.
The most commonly reported device issues were cracks (8 cases), followed by twisted or bent material (6 cases), and deformation or stretching of material (5 cases). These problems highlight concerns about the device’s structural integrity during use. Other issues, such as physical resistance, fluid or blood leaks, and retraction difficulties, were less frequent but still notable. For details on all reported device events, see Table 1 & Fig. 2.
The primary indication for device use was middle cerebral artery (MCA) stroke, with nine reported cases, followed by internal carotid artery (ICA) stroke with three cases. Three additional reports only specified stroke without indicating the type. Less common were basilar artery strokes (2 cases) and vertebral artery strokes (1 case). Some reports involved combined stroke presentations, including one case involving both the MCA and anterior cerebral artery (ACA), and another involving the MCA and ICA.
Unfortunately, the reports lacked demographic data, with only three reports including patient age and gender. Physicians submitted nineteen reports, and only 12 catheters were returned to the manufacturer for further analysis.

DISCUSSION

Main findings and significance

Based on reports from the FDA MAUDE database, this study is the first to analyze post-market safety signals related to the recently introduced Cerenovus CEREGLIDE 71 Intermediate Catheter. While thromboembolism, intracranial hemorrhage, and arterial injury were the most common patient-related complications, structural failures such as cracks, deformation, stretching, and retraction issues were the most frequently reported device-related problems. Although the catheter has advanced navigational capabilities, its practical use reveals a risk profile that warrants further research. Because they provide early safety signals for a new thrombectomy device, these findings are clinically important and will assist regulatory agencies, device designers, and clinicians in making informed decisions.

Comparison with existing literature

Our findings are consistent with an expanding body of research on the risks associated with large-bore aspiration catheters used in mechanical thrombectomy. Previous reports of structural failures in a variety of devices have included coil separation, kinking, and catheter breakage [1,8]. Patients with stenotic or tortuous vasculature, where device stress is increased, should be especially concerned about these mechanical problems. In fact, ICA tortuosity has been linked to increased hemorrhagic risks, longer procedure times, and a lower first-pass effect [3,11]. According to our data, these anatomical difficulties make it more difficult to advance and retrieve a catheter, which raises the possibility of vascular trauma and device failure.
The observed vasospasm and thromboembolic events could be caused by endothelial irritation or catheter-induced clot fragmentation, both of which have been previously reported in previous research [17,26]. Despite the fact that ADAPT has been linked to lower symptomatic ICH rates and faster reperfusion than stent retrievers [10,23], the device-related thromboembolism in our series might be a reflection of actual handling or procedure errors.
Our results are in line with previous meta-analyses that found several contributing factors, such as multiple device passes, distal occlusions, and intraprocedural complications, and that showed symptomatic ICH rates of 7-8% in relation to hemorrhagic complications [9,15]. Our findings highlight that inappropriate catheter manipulation or undetected vessel pathology may still result in vessel rupture or dissection, despite some research suggesting that aspiration-based techniques lower the risk of sICH [10].
It is also critical to compare our findings with the encouraging preliminary results reported in the only available literature on real-world use of the CEREGLIDE 71—a multi-center retrospective study by Majidi et al., presented as a conference abstract. Despite 81.7% of cases involving challenging anatomies, the catheter achieved a 99.1% navigation success rate. Reperfusion outcomes included a 35% first-pass mTICI ≥2c and a 95% final mTICI ≥2b, with direct aspiration alone effective in 60% of patients. The safety profile was favorable, with no device-related adverse events and low rates of symptomatic intracranial hemorrhage (3.3%) and vessel dissection (1.7%) [18]. The catheter is also set to be included in the upcoming phase of the CERENOVUS EXCELLENT Registry. In contrast to the varied and uncontrolled scenarios represented in the MAUDE database, the Majidi study reflects controlled institutional settings, typically involving experienced operators and selective patient inclusion.

Study implications

This study raises awareness of potential complications beyond initial trial environments and provides timely insights into the CEREGLIDE 71 catheter’s real-world safety profile. It emphasises the value of procedural vigilance for interventionalists, particularly when managing resistance or navigating complex anatomy. These findings might also be helpful to device makers in improving the flexibility and durability of designs for upcoming catheter generations. Furthermore, by showing actual complication patterns, our findings can help guide informed consent conversations, particularly with high-risk patients.
This study supports the need for post-market surveillance systems to supplement pre-market assessments, which might miss uncommon or delayed device-related risks, from a policy perspective. MAUDE is a crucial but underutilised tool for early adverse event detection, particularly during the launch phase of new devices.

Limitations and strengths

The MAUDE database’s inherent limitations, such as clinicians’ voluntary reporting, unstructured or incomplete event narratives, the absence of denominator data (such as the number of devices used), and possible duplication, all affect our analysis [6,32]. Therefore, it is not appropriate to use our results to compare complication rates across devices or to interpret them as incidence rates.
The study does, however, have some significant advantages. It offers an early real-world safety profile for a recently released device and is the first to systematically analyse adverse events related to CEREGLIDE 71. It is a useful resource for academic and clinical audiences due to its systematic review, categorisation of adverse events, and contextualisation with other literature. Furthermore, by providing a more comprehensive perspective on device performance, the study supports the results of real-world registries (like the EXCELLENT registry) and early clinical trials.

Future directions

Several future directions should be considered to enhance device safety evaluation. First, data consistency, completeness, and comparability across reports could be improved by establishing standardized reporting templates within the MAUDE database. Strengthening surveillance and providing more insightful cross-device analysis may also be supported by an improved adverse event reporting structure. Second, researchers could combine structured, prospective outcomes with passive surveillance by integrating MAUDE reports with real-world registry data, such as the EXCELLENT and COMPASS registries. This integration would offer a more comprehensive view of device safety and efficacy. Third, extensive real-world datasets should be utilized to develop predictive models for device failure, particularly in cases involving tortuous vascular anatomy. These models could assist physicians in risk assessment and pre-procedural planning. Lastly, creating prospective, multicenter registries focused exclusively on large-bore aspiration catheters is justified. To better evaluate device performance across diverse clinical settings, these registries should stratify patients based on procedural variations, operator experience, and anatomical complexity. Long-term outcome studies and ongoing safety monitoring will be essential as large-bore aspiration technologies evolve, ensuring patient safety and supporting innovation.

CONCLUSIONS

The Cereglide 71 catheter, when used in ADAPT techniques, offers additional advantages for mechanical thrombectomy, although it is linked to various complications. Recognizing these complications by reviewing the reports documented in the FDA MAUDE database boosts awareness of their potential occurrences. This understanding makes it essential to adopt specific surgical techniques to reduce risks.

NOTES

Disclosures

The authors report no conflict of interest concerning the materials or methods used in this study or the findings specified in this paper.

Fig. 1.
Flowchart illustrating the identification and selection process of MAUDE reports related to the Cerenovus CEREGLIDE 71 Intermediate Catheter.
jcen-2025-e2025-01-004f1.jpg
Fig. 2.
Distribution of reported device problems associated with the Cereglide 71 Catheter based on MAUDE database analysis
jcen-2025-e2025-01-004f2.jpg
Table 1.
Frequency of reported device problems and malfunctions in adverse event reports
Device problem Occurrences Device problem Occurrences
Crack 8 Component or accessory incompatibility 2
Material Twisted/Bent 6 Physical resistance/Sticking 2
Stretched 5 Kinked upon removal from the package 1
Material deformation 5 Difficult to remove 1
Retraction problem 4 Deformation due to compressive stress 1
Difficult to advance 3 Material too soft/Flexible 1
Fluid/Blood leak 2 Peeled/Delaminated 1
Material puncture/Hole 2 Detachment of device or device component 1
Material separation 2

REFERENCES

1. Akaike N, Ikeda H, Uezato M, Yamashita H, Kinosada M, Kurosaki Y, et al. Disconnection of a stent retriever’s pushwire caught by an accordion-like deformed aspiration catheter during mechanical thrombectomy: Illustrative case. J Neurosurg Case Lessons. 2023 Jun;5(26):CASE23147.
crossref pmid pmc
2. Batheja A, Bentan M, Coelho DH. Considerations for use of the MAUDE database in otolaryngology-head & neck surgery research. Otolaryngol Head Neck Surg. 2024 Jul;171(1):320-2.
crossref pmid
3. Bilgin C, Gupta R, Orscelik A, Hassankhani A, Senol YC, Kobeissi H, et al. Impact of carotid artery tortuosity on mechanical thrombectomy outcomes: A systematic review. Neuroradiol J. 2025 Feb;19714009251317500.
crossref pdf
4. Boisseau W, Escalard S, Fahed R, Lapergue B, Smajda S, Maier B, et al. Direct aspiration stroke thrombectomy: A comprehensive review. J Neurointerv Surg. 2020 Nov;12(11):1099-106.
crossref pmid
5. Case BC, Forrestal BJ, Yerasi C, Khan JM, Khalid N, Shlofmitz E, et al. Real-world experience of the sentinel cerebral protection device: Insights from the FDA Manufacturer and User Facility Device Experience (MAUDE) database. Cardiovasc Revasc Med. 2020 Feb;21(2):235-8.
crossref pmid
6. Davies C, Nieri CA, Sheyn A, Rangarajan S, Yawn RJ. The use and utility of food and drug administration adverse event data from the manufacturer and user facility device experience database in otology: A systematic review. Otol Neurotol. 2023 Jul;44(6):534-41.
crossref pmid
7. Ding Q, Liu S, Yao Y, Liu H, Cai T, Han L. Global, regional, and national burden of ischemic stroke, 1990-2019. Neurology. 2022 Jan;98(3):e279-90.
crossref pmid
8. Dolia JN, Mohammaden MH, Tarek MA, Damiani M, Grossberg JA, Pabaney A, et al. Structural analysis of aspiration catheters and procedural outcomes: An analysis of the SVIN registry. SVIN. 2024 Jul;4(4).
crossref
9. Hao Z, Yang C, Xiang L, Wu B, Liu M. Risk factors for intracranial hemorrhage after mechanical thrombectomy: A systematic review and meta-analysis. Expert Rev Neurother. 2019 Oct;19(10):927-35.
crossref pmid
10. Ironside N, Chen CJ, Chalhoub RM, Wludyka P, Kellogg RT, Al Kasab S, et al. Risk factors and predictors of intracranial hemorrhage after mechanical thrombectomy in acute ischemic stroke: Insights from the Stroke Thrombectomy and Aneurysm Registry (STAR). J Neurointerv Surg. 2023 Nov;15(e2):e312-22.
crossref pmid pmc
11. Koge J, Tanaka K, Yoshimoto T, Shiozawa M, Kushi Y, Ohta T, et al. Internal carotid artery tortuosity: Impact on mechanical thrombectomy. Stroke. 2022 Aug;53(8):2458-67.
crossref pmid pmc
12. Kortman VG, Verberne Y, Atalla MA, Sakes A. Advancements in aspiration catheter tip design for thrombectomy: A comprehensive patent review. Front Med Technol. 2024 Nov;6:1388638.
crossref pmid pmc
13. Krishnamurthi RV, Ikeda T, Feigin VL. Global, Regional and Country-Specific Burden of Ischaemic Stroke, Intracerebral Haemorrhage and Subarachnoid Haemorrhage: A Systematic Analysis of the Global Burden of Disease Study 2017. Neuroepidemiology. 2020 Feb;54(2):171-9.
crossref pmid pdf
14. Lapergue B, Blanc R, Gory B, Labreuche J, Duhamel A, Marnat G. Effect of endovascular contact aspiration vs stent retriever on revascularization in patients with acute ischemic stroke and large vessel occlusion. The ASTER randomized clinical trial. JAMA. 2017 Aug;318(5):443-52.
crossref pmid pmc
15. Lee H, Qureshi AM, Mueller-Kronast NH, Zaidat OO, Froehler MT, Liebeskind DS, et al. Subarachnoid hemorrhage in mechanical thrombectomy for acute ischemic stroke: Analysis of the STRATIS registry, systematic review, and meta-analysis. Front Neurol. 2021 May;12:663058.
crossref pmid pmc
16. Li J, Ribo M. REACT aspiration catheters: Clinical experience and technical considerations. Neurointervention. 2022 Jul;17(2):70-7.
crossref pmid pmc pdf
17. Liu Y, Gebrezgiabhier D, Reddy AS, Davis E, Zheng Y, Arturo Larco JL, et al. Failure modes and effects analysis of mechanical thrombectomy for stroke discovered in human brains. J Neurosurg. 2022 Jan;136(1):197-204.
crossref pmid
18. Majidi S, Grandhi R, Samaniego E, Kilburg CJ, Budohoski KP, Reddi P, et al. CEREGLIDE 71 Aspiration Catheter for stroke thrombectomy: An initial multicenter US experience. Stroke: Vascular and Interventional Neurology. 2024 Nov;4(Suppl_1):Abstract 020.
crossref
19. Marnat G, Barreau X, Detraz L, Bourcier R, Gory B, Sgreccia A, et al. First-line sofia aspiration thrombectomy approach within the endovascular treatment of ischemic stroke multicentric registry: Efficacy, safety, and predictive factors of success. AJNR Am J Neuroradiol. 2019 Jun;40(6):1006-12.
crossref pmid pmc
20. McMeekin P, Flynn D, James M, Price CI, Ford GA, White P. Updating estimates of the number of UK stroke patients eligible for endovascular thrombectomy: Incorporating recent evidence to facilitate service planning. Eur Stroke J. 2021 Dec;6(4):349-56.
crossref pmid pmc pdf
21. Phan K, Dmytriw AA, Teng I, Moore JM, Griessenauer C, Ogilvy C, et al. A direct aspiration first pass technique vs standard endovascular therapy for acute stroke: A systematic review and meta-analysis. Neurosurgery. 2018 Jul;83(1):19-28.
crossref pmid pdf
22. Rai AT, Seldon AE, Boo S, Link PS, Domico JR, Tarabishy AR, et al. A population-based incidence of acute large vessel occlusions and thrombectomy eligible patients indicates significant potential for growth of endovascular stroke therapy in the USA. J Neurointerv Surg. 2017 Aug;9(8):722-6.
crossref pmid pmc
23. Savastano L, Gebrezgiabhier D, Arturo Larco J, Madhani S, Shahid A, Liu Y. E-015 Failure mechanisms of current thrombectomy devices identified in a human brain model: Iatrogenic embolization, residual and recurrent large vessel occlusion, persistent perforating artery occlusion, and arterial collapse, traction and avulsion. Journal of NeuroInterventional Surgery. 2021;13:A71.
crossref
24. Schartz D, Ellens N, Kohli GS, Rahmani R, Akkipeddi SMK, Colby GP, et al. Impact of aspiration catheter size on clinical outcomes in aspiration thrombectomy. J Neurointerv Surg. 2023 Sep;15(e1):e111-6.
crossref pmid
25. Sila D, Lenski M, Vojtková M, Elgharbawy M, Charvát F, Rath S. Efficacy of mechanical thrombectomy using penumbra ACETM aspiration catheter compared to Stent Retriever SolitaireTM FR in patients with acute ischemic stroke. Brain Sci. 2021 Apr;11(4):504.
crossref pmid pmc
26. Stapleton CJ, Leslie-Mazwi TM, Torok CM, Hakimelahi R, Hirsch JA, Yoo AJ, et al. A direct aspiration first-pass technique vs stentriever thrombectomy in emergent large vessel intracranial occlusions. J Neurosurg. 2018 Feb;128(2):567-74.
crossref pmid
27. Tortora R, Flora G, Frauenfelder G, Locatelli G, Siani A, Saponiero R, et al. P031/333 A very rare complication during mechanical thrombectomy: The breakage of the aspiration catheter’s tip and his rescue. Journal of NeuroInterventional Surgery. 2023;15:A32.
crossref
28. Tsang COA, Cheung IHW, Lau KK, Brinjikji W, Kallmes DF, Krings T. Outcomes of stent retriever versus aspiration-first thrombectomy in ischemic stroke: A systematic review and meta-analysis. AJNR Am J Neuroradiol. 2018 Nov;39(11):2070-6.
crossref pmid pmc
29. Turc G, Kim AS. First-line use of contact aspiration or stent retriever thrombectomy for large vessel occlusion stroke. Stroke. 2019 Sep;50(9):2634-6.
crossref pmid
30. Vargas J, Blalock J, Venkatraman A, Anagnostakou V, King RM, Ewing JA, et al. Efficacy of beveled tip aspiration catheter in mechanical thrombectomy for acute ischemic stroke. J Neurointerv Surg. 2021 Sep;13(9):823-6.
crossref pmid pmc
31. Zhu W, He X, Huang D, Jiang Y, Hong W, Ke S, et al. Global and regional burden of ischemic stroke disease from 1990 to 2021: An age-period cohort analysis. medRxiv preprint. 2024 Aug.
crossref
32. Ziapour B, Zaepfel C, Iafrati MD, Suarez LB, Salehi P. A systematic review of the quality of cardiovascular surgery studies that extracted data from the MAUDE database. J Vasc Surg. 2021 Nov;74(5):1708-20.e5.
crossref pmid
TOOLS
METRICS Graph View
  • 1 Crossref
  •  0 Scopus
  • 1,173 View
  • 37 Download
ORCID iDs

Ibraheem Alkhawaldeh
https://orcid.org/0000-0002-0187-1583

Related articles


ABOUT
BROWSE ARTICLES
EDITORIAL POLICY
FOR CONTRIBUTORS
Editorial Office
The Journal of Cerebrovascular and Endovascular Neurosurgery (JCEN), Department of Neurosurgery, Wonkwang University
School of Medicine and Hospital, 895, Muwang-ro, Iksan-si, Jeollabuk-do 54538, Korea
Tel: +82-2-2279-9560    Fax: +82-2-2279-9561    E-mail: editor.jcen@the-jcen.org                

Copyright © 2026 by Korean Society of Cerebrovascular Surgeons and Korean NeuroEndovascular Society.

Developed in M2PI

Close layer
prev next